Alphenix INFX-8000F, interventional fluoroscopic x-ray system

Agency
U.S. Food and Drug Administration
Recall number
Z-1177-2026
Category
Medical devices
Recall date
2025-12-19
Enforcement report date
2026-02-04
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Canon Medical System, USA, INC.

From the FDA Enforcement Report dated 2026-02-04. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Quantity
98 units
Distribution
Worldwide distribution - US Nationwide and the country of Dominican Republic.
Code information
serial numbers: W1C0582204, B2A1582208, W2A1052118, A3582035, A2A2362689, B2B18Y2813, W1D07Y2447, B4582113, W2A1042111, W1C0622270, A3592048, W4A11X2138, W2A09Y2060, B2A16Z2476, W2A09Y2066, W4A1122041, B2A1482032, B2A1482027, A2A2252435, W1D0722369, B2B15Y2247, W1B0542168, B2A1492049, W1D0722371, A2A2032031, A1C1972103, W4C1352375, B2B1862737, W1D0842499, W4A1212179, W4A11Z2171, W4A11Z2173, W1C0652298, B2A1692419, B2A1552170, A2A21Y2350, W4A1242226, B2B1952848, A1C19Z2180, B2A1662371, W1D0892555, W1D0822476, W1D06Z2356, B2A1682394, B2A17Y2619, B2A17Y2624, W1D0762406, A2A2172283, W4A11Y2155, A2A2272460, W1D0832487, B2B1822667, A1C1992146, W4A1112031, W4C1382409, A2A2292508, W4A13X2015, W4A1192130, B2A1712510, B2A1762562, W4A1142066, W1C0622274, A4552098, W4A12X2298, W1D0812460, A4542088, W1D0712363, A2A21X2348, W1C0552179, W4A1142065, B2A17Z2626, W4D13Y2444, W2A1042110, A2A20Z2159, W4A11Y2147, W1C0642289, B2B1852720, B2A1592219, W4B13Y2016, B2B1922843, B2B1822668, W4B12Y2312, A2A2052056, A2A2062062, A2A2072075, A2A2042041, A1C19Z2181, W2A1092163, A2A2012005, A2A20Y2147, A1C19Y2161, B2A1612289, W4A1112030, W2A0982042, B2B18X2802, B2A1712503, B2A1712499, B2A14Z2084.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands