Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1201-2026
Category
Medical devices
Recall date
2025-12-24
Enforcement report date
2026-02-04
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Zimmer Surgical Inc

From the FDA Enforcement Report dated 2026-02-04. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A
Hazard / reason
Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Quantity
206
Distribution
International distribution to the countries of Canada and EMEA only.
Code information
Lot Code: Model No. 60500010100 ; UDI (01)00889024626454(11)240410(10)66722609(21)50066722609001; Lot No. 66722609 ; Serial No. 50066722609001 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609002; Lot No. 66722609 ; Serial No. 50066722609002 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609003" ; Lot No. 66722609 ; Serial No. 50066722609003 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609004" ; Lot No. 66722609 ; Serial No. 50066722609004 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609005" ; Lot No. 66722609 ; Serial No. 50066722609005 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610001" ; Lot No. 66722610 ; Serial No. 50066722610001 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610002" ; Lot No. 66722610 ; Serial No. 50066722610002 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610003" ; Lot No. 66722610 ; Serial No. 50066722610003 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610004" ; Lot No. 66722610 ; Serial No. 50066722610004 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610005" ; Lot No. 66722610 ; Serial No. 50066722610005 Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611001" ; Lot No. 66722611 ; Serial No.…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Possible model numbers (as listed; not verified as model numbers)
  • 60500010100

Brands