SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1216-2026
- Category
- Medical devices
- Recall date
- 2025-12-03
- Enforcement report date
- 2026-02-04
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Philips North America
From the FDA Enforcement Report dated 2026-02-04. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1216-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.
- Hazard / reason
- The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
- Quantity
- 3 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.
- Code information
- MR systems with SW version R11.1 and R12.1. Product Code (REF): 781270. UDI-DI: 00884838095083. Serial Numbers: 24066, 38113, 17356.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II