Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1223-2026
- Category
- Medical devices
- Recall date
- 2024-02-17
- Enforcement report date
- 2026-02-11
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Olympus Corporation of the Americas
From the FDA Enforcement Report dated 2026-02-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1223-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
- Hazard / reason
- A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
- Quantity
- 3361 units
- Distribution
- US Nationwide distribution.
- Code information
- Model/Catalog Number: KD-645L UDI: 04953170407857 Lot Number(s): 2ZK, 31K, 32K, 33K, 35K, 3XK, 3YK, 3ZK, 42K, 43K, 44K, 45K, 46K, 47K, 48K, 49K, 4XK, 4YK, 4ZK, 51K, 52K, 53K, 54K, 55K, 56K, 57K, 58K, 59K
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 4953170407857