CellMek SPS Sample Preparation System, REF: C44603, with software

Agency
U.S. Food and Drug Administration
Recall number
Z-1235-2026
Category
Medical devices
Recall date
2025-12-23
Enforcement report date
2026-02-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Beckman Coulter, Inc.

From the FDA Enforcement Report dated 2026-02-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Quantity
50
Distribution
Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.
Code information
UDI-DI: 15099590750312. Software Version 2.3.90.0
Recall type
Voluntary: Firm initiated
Product type
Devices