Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1302-2026
- Category
- Medical devices
- Recall date
- 2025-12-11
- Enforcement report date
- 2026-02-11
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Fujirebio Diagnostics, Inc.
From the FDA Enforcement Report dated 2026-02-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1302-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
- Hazard / reason
- Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
- Quantity
- 1569 units
- Distribution
- US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
- Code information
- Model/Catalog Number: 81550; UDI-DI: 05414470815500; Lot Numbers: D4C6028U, D4C6029U, D4C6051U, D4C6052U;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 5414470815500