Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1311-2026
Category
Medical devices
Recall date
2025-11-17
Enforcement report date
2026-02-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
ETAC A/S

From the FDA Enforcement Report dated 2026-02-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.
Hazard / reason
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Quantity
25 units
Distribution
Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
Code information
(1) Model/Catalog Number: 1430116; UDI-DI: TBD; (2) Model/Catalog Number: 1430117; UDI-DI: TBD; (3) Model/Catalog Number: 1430115; UDI-DI: TBD;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands