Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1321-2026
- Category
- Medical devices
- Recall date
- 2025-12-22
- Enforcement report date
- 2026-02-18
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-02-18. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1321-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH;
- Hazard / reason
- These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
- Quantity
- 41
- Distribution
- Distribution US nationwide.
- Code information
- 401582RH UDI-DI (case) 20888277407517¿ (ea) 10888277407510 Lots EP250618 EP250630; 401940RH UDI-DI 10888277407480 Lot EP250618; 401904RH UDI-DI (case) 20888277407609¿ (ea) 10888277407602 Lots EP250618 EP250630; 401914RH UDI-DI (case) 20888277407524¿ (ea) 10888277407527 Lot EP250618; 401932RH UDI-DI (case) 20888277407678¿ (ea) 10888277407671 Lots EP250618 EP250630; 401652RH UDI-DI 10888277407572 Lots EP250618 EP250630
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I
Identifiers
- Barcodes (UPC/GTIN)
- 10888277407480
- 10888277407572