Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1324-2026
Category
Medical devices
Recall date
2025-12-22
Enforcement report date
2026-02-18
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-02-18. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on Siemens Systems 10438577RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter, Use on Siemens Systems 10439011RH
Hazard / reason
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Quantity
94
Distribution
Distribution US nationwide.
Code information
10439072RH UDI-DI 10197344157297 Lots EP250724 EP250808; 10439236RH UDI-DI 10197344157303 Lots EP250724 EP250808; 10438577RH UDI-DI 10197344157273 Lots EP250724 EP250808; 10439011RH UDI-DI (case) 20197344157287¿ (ea) 10197344157280¿ Lots EP250724 EP250808
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 10197344157297
  • 10197344157303
  • 10197344157273

Brands