Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1384-2026
- Category
- Medical devices
- Recall date
- 2026-01-07
- Enforcement report date
- 2026-02-25
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-02-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1384-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A
- Hazard / reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Quantity
- 551 units
- Distribution
- Worldwide distribution - US Nationwide.
- Code information
- 1) DYNJ0185290G, UDI-DI: 10193489378054(each), 40193489378055(case), Lot Number: 22LBQ336; 2) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 24BBN429; 3) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23KBI089; 4) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23HBG596; 5) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBL644; 6) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBG152; 7) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23EBH550; 8) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23CBQ546; 9) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22NBF780; 10) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22LBC300; 11) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBP453; 12) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBM263; 13) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBF686; 14) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22CBY841; 15) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBR457; 16) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBA253; 17) DYNJ905567A, UDI-DI: 10193489348927(each),…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10193489378054
- 10193489348927
- Model numbers
- DYNJ0185290G
- DYNJ905567A