Medline Convenience Kits: 1) PPE KIT, Model Number: DYK1011945P
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1408-2026
- Category
- Medical devices
- Recall date
- 2026-01-07
- Enforcement report date
- 2026-02-25
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-02-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1408-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Convenience Kits: 1) PPE KIT, Model Number: DYK1011945P
- Hazard / reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Quantity
- 7575 units
- Distribution
- Worldwide distribution - US Nationwide.
- Code information
- 1) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 25IFA006; 2) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 24GFA027; 3) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 24FFA115; 4) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 24BFA115; 5) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23LFA050; 6) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23GFA079; 7) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23EFA070; 8) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23EFA007; 9) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23DFA024; 10) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23CFA070; 11) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23BFA124; 12) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23BFA006; 13) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23AFA090; 14) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 23AFA049; 15) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 22LFA068; 16) DYK1011945P, UDI-DI: 10884389389973(each), 40884389389974(case), Lot Number: 22JFA111; 17) DYK1011945P, UDI-DI: 10884389389973(each),…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10884389389973
- Model numbers
- DYK1011945P