Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1411-2026
- Category
- Medical devices
- Recall date
- 2026-01-07
- Enforcement report date
- 2026-02-25
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-02-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1411-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
- Hazard / reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Quantity
- 9720 units
- Distribution
- Worldwide distribution - US Nationwide.
- Code information
- 1) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2025051290; 2) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2025031190; 3) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2023062390; 4) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2023101290; 5) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2022030390; 6) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2022072590; 7) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020042290; 8) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020052290; 9) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020072290; 10) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020082690; 11) DT8670A, UDI-DI: 00653160177473(each), 10653160177470(case), Lot Number: 2020051390
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10653160345251
- 10653160327752
- 653160177473
- Model numbers
- DM1035A
- DM1035
- DT8670A