Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1464-2026
- Category
- Medical devices
- Recall date
- 2026-01-08
- Enforcement report date
- 2026-03-04
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-03-04. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1464-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
- Hazard / reason
- Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
- Quantity
- 2020
- Distribution
- US Nationwide distribution.
- Code information
- UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II