EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Agency
U.S. Food and Drug Administration
Recall number
Z-1480-2026
Category
Medical devices
Recall date
2026-01-21
Enforcement report date
2026-03-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Staar Surgical AG

From the FDA Enforcement Report dated 2026-03-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
Quantity
7
Distribution
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Code information
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 840311304363
Model numbers
  • VTICM5

Brands