EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1480-2026
- Category
- Medical devices
- Recall date
- 2026-01-21
- Enforcement report date
- 2026-03-11
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Staar Surgical AG
From the FDA Enforcement Report dated 2026-03-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1480-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
- Quantity
- 7
- Distribution
- International distribution to the countries of India, Iran, Korea, Saudi Arabia.
- Code information
- UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 840311304363
- Model numbers
- VTICM5