Atellica CH Urine Albumin (UAlb). Material Number: 11537225

Agency
U.S. Food and Drug Administration
Recall number
Z-1484-2026
Category
Medical devices
Recall date
2026-01-20
Enforcement report date
2026-03-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Siemens Healthcare Diagnostics, Inc.

From the FDA Enforcement Report dated 2026-03-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Quantity
4,885 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
Code information
Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 630414611099

Brands