Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1494-2026
- Category
- Medical devices
- Recall date
- 2026-01-08
- Enforcement report date
- 2026-03-11
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Olympus Corporation of the Americas
From the FDA Enforcement Report dated 2026-03-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1494-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.
- Hazard / reason
- Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
- Quantity
- 55 units (All OUS)
- Distribution
- International distribution in the country of Canada, Germany, Singapore, Australia, and India.
- Code information
- Model/Catalog Number: SPL-SR, contains generator SPL-G; UDI: SPL-SR 00821925043824, SPL-G 00821925044203; Serial Numbers: CG5026, CG5027, CG5028, CG5029, CG5030, CG5031, CG5032, CG5033, CG5034, CG5035, CG5036, CG5037, CG5039, CG5040, CG5041, CG5042, CG5043, CG5045, CG5046, CG5047, CG5048, CG5049, CG5050, CG5051, CG5052, CG5053, CG5054, CG5055, CG5056, CG5057, CG5058, CG5059, CG5060, CG5061, CG5062, CG5063, CG5065, CG5066, CG5067, CG6037, CG6061, CG6069, CG6070, CG6071, CG6088, CG6094, CG6095, CG6096, CG6097, CG6098, CG6099, CG7000, CG7001, CG7002, CG7003;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II