IBA Proton Therapy System - PROTEUS 235

Agency
U.S. Food and Drug Administration
Recall number
Z-1497-2026
Category
Medical devices
Recall date
2026-02-03
Enforcement report date
2026-03-18
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Ion Beam Applications S.A.

From the FDA Enforcement Report dated 2026-03-18. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Quantity
4 units (1 US, 3 OUS)
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Code information
PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
Recall type
Voluntary: Firm initiated
Product type
Devices