Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1571-2026
Category
Medical devices
Recall date
2026-01-30
Enforcement report date
2026-03-25
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
GE Medical Systems, LLC

From the FDA Enforcement Report dated 2026-03-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Centricity Universal Viewer Software Versions 5.0 SP6 through UV 5.0 SP7.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-026, 2088026-115, 2088026-043, K2042VJED, 2088026-132, 2088026-003, 2088026-110, 2089629-003, 2066908-136, 2066908-150
Hazard / reason
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Quantity
40 units
Distribution
Worldwide - US Nationwide distribution.
Code information
Product ID Numbers: 4446-2-Centricity Universal Viewer-00558643 100256-3-Centricity Universal Viewer-00558700 101044-1-Centricity Universal Viewer-00558694 13240-2-Centricity Universal Viewer-00558103 4441-1-Centricity Universal Viewer-00558677 4578-1-Centricity Universal Viewer-00558340 10831-2-Centricity Universal Viewer-00558682 656341-Centricity Universal Viewer-01431499 50567-Centricity Universal Viewer-02176369 2690973-Centricity Universal Viewer-01454622 EG1526-Centricity Universal Viewer-00158976 A91901-Centricity Universal Viewer-00044549 M3488569-Centricity Universal Viewer-00071707 25540620 - HC4105-Centricity Universal Viewer-00001393 416005-Centricity Universal Viewer-00260808 1868708X - 556162-Centricity Universal Viewer-00157857 HKOB01-Centricity Universal Viewer-01282268 BMH-Centricity Universal Viewer-01820534 JIPMER-Centricity Universal Viewer-01820555 4347967-Centricity Universal Viewer-01474540 651342-Centricity Universal Viewer-01365218 651342-Centricity Universal Viewer-01365213 A3431341-Centricity Universal Viewer-00545677 473265-Centricity Universal Viewer-01442014 JP-131575-Centricity Universal Viewer-01275307 442900-Centricity Universal Viewer-01460668 577127-Centricity Universal Viewer-01444349 349551-Centricity Universal Viewer-02350518 PL3277-Centricity Universal Viewer-00038831 RU3866-Centricity Universal Viewer-01445731 RU8945-Centricity Universal Viewer-01445740 RU2580-Centricity Universal Viewer-01445743 1354732-Centricity Universal…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Model numbers
  • K2042VJED
Possible model numbers (as listed; not verified as model numbers)
  • 2088026-026
  • 2088026-115
  • 2088026-043
  • 2088026-132
  • 2088026-003
  • 2088026-110
  • 2089629-003
  • 2066908-136
  • 2066908-150