MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Agency
U.S. Food and Drug Administration
Recall number
Z-1572-2026
Category
Medical devices
Recall date
2026-02-02
Enforcement report date
2026-03-25
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Howmedica Osteonics Corp.

From the FDA Enforcement Report dated 2026-03-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Hazard / reason
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Quantity
1098 units
Distribution
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Code information
Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II