Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: in…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1585-2026
Category
Medical devices
Recall date
2026-02-05
Enforcement report date
2026-03-25
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Diagnostica Stago, Inc.

From the FDA Enforcement Report dated 2026-03-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: in vitro diagnostic. The STA - Liatest D-Di kit is an assay for the quantitative determination of D-dimer in venous plasma for use on STA-R, STA Compact and STA Satellite analyzers by professional laboratory personnel.
Hazard / reason
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
Quantity
34000 units (28446 US, 5554 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Singapore, Guam, Japan, South Korea, and Russia.
Code information
Model/Catalog Number: REF 00515; UDIs: (1) (01)03607450005158(11)240630(17)250930(10)273015(241)00515, (2) (01)03607450005158(11)250325(17)260630(10)273127(241)00515, (3)(01)03607450005158(11)250416(17)260731(10)273275(241)00515, (4)(01)03607450005158(11)250527(17)260831(10)273567(241)00515, (5)(01)03607450005158(11)250618(17)260930(10)273681(241)00515, (6)(01)03607450005158(11)250716(17)261031(10)273808(241)00515, (7)(01)03607450005158(11)250716(17)261231(10)273808(241)00515, (8)(01)03607450005158(11)250909(17)261231(10)274108(241)00515 (9) (01)03607450005158(11)250326(17)260630(10)273128(241)00515, (10) (01)03607450005158(11)250429(17)260731(10)273606(241)00515, (11) (01)03607450005158(11)250709(17)261031(10)273805(241)00515, (12) (01)03607450005158(11)251007(17)270131(10)274219(241)00515; Lot numbers: (1)273015, (2)273127, (3)273275, (4)273567, (5)273681, (6)273808, (7)273995, (8)274108, (9)273128, (10)273606, (11)273805, (12)274219;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II