Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: in…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1585-2026
- Category
- Medical devices
- Recall date
- 2026-02-05
- Enforcement report date
- 2026-03-25
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Diagnostica Stago, Inc.
From the FDA Enforcement Report dated 2026-03-25. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1585-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: in vitro diagnostic. The STA - Liatest D-Di kit is an assay for the quantitative determination of D-dimer in venous plasma for use on STA-R, STA Compact and STA Satellite analyzers by professional laboratory personnel.
- Hazard / reason
- After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
- Quantity
- 34000 units (28446 US, 5554 OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Singapore, Guam, Japan, South Korea, and Russia.
- Code information
- Model/Catalog Number: REF 00515; UDIs: (1) (01)03607450005158(11)240630(17)250930(10)273015(241)00515, (2) (01)03607450005158(11)250325(17)260630(10)273127(241)00515, (3)(01)03607450005158(11)250416(17)260731(10)273275(241)00515, (4)(01)03607450005158(11)250527(17)260831(10)273567(241)00515, (5)(01)03607450005158(11)250618(17)260930(10)273681(241)00515, (6)(01)03607450005158(11)250716(17)261031(10)273808(241)00515, (7)(01)03607450005158(11)250716(17)261231(10)273808(241)00515, (8)(01)03607450005158(11)250909(17)261231(10)274108(241)00515 (9) (01)03607450005158(11)250326(17)260630(10)273128(241)00515, (10) (01)03607450005158(11)250429(17)260731(10)273606(241)00515, (11) (01)03607450005158(11)250709(17)261031(10)273805(241)00515, (12) (01)03607450005158(11)251007(17)270131(10)274219(241)00515; Lot numbers: (1)273015, (2)273127, (3)273275, (4)273567, (5)273681, (6)273808, (7)273995, (8)274108, (9)273128, (10)273606, (11)273805, (12)274219;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II