I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1603-2026
- Category
- Medical devices
- Recall date
- 2026-01-23
- Enforcement report date
- 2026-04-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- I.T.S. GmbH
From the FDA Enforcement Report dated 2026-04-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1603-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral Tibia Locking) Proximal Lateral Tibia Plate; Article Numbers: 21321-7, 21321-12, 21322-7, 21322-12. 3.PTL (Proximal Lateral Tibia Locking) Proximal Medial Tibia Plate; Article Numbers: 21325-4, 21325-7.
- Hazard / reason
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Quantity
- 176 units
- Distribution
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Code information
- 1. Article Number (Lot Numbers): 21301-5 (26/F62569), 21301-7 (26/F62570), 21301-9 (26/F67518), 21301-13 (26/F62568, 26/F67815), 21302-5 (26/F67520), 21302-7 (26/F67821), 21302-9 (26/F62572, 26/F67822), 21302-13 (26/F62571, 26/F67519, 26/F67820). 2. Article Number (Lot Numbers): 21321-7 (26/F54955, 26/F62576), 21321-12 (26/F55883), 21322-7 (26/F55890, 26/F62578), 21322-12 (26/F62577). 3. Article Number (Lot Numbers): 21325-4 (26/F70528, 679/022433, 679/032217, 679/102240, 679/102242, 679/102303), 21325-7 (679/102241, 679/102304).
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II