Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1631-2026
- Category
- Medical devices
- Recall date
- 2026-03-04
- Enforcement report date
- 2026-04-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Koven Technology, Inc.
From the FDA Enforcement Report dated 2026-04-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1631-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.
- Hazard / reason
- Product labeling includes a fetal indication for use that is not cleared under its 510(k).
- Quantity
- 34
- Distribution
- US Nationwide distribution in the states of LA, NC, MT, TX MA, OR, NY, IL, CO, GA, FL, WA, WI, MO, AZ, IN.
- Code information
- Lot Code: BIDOP 3 #00894912002043 BT2M20S8C #00894912002326
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II