Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1637-2026
Category
Medical devices
Recall date
2026-02-18
Enforcement report date
2026-04-08
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Baxter Healthcare Corporation

From the FDA Enforcement Report dated 2026-04-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.
Hazard / reason
Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
Quantity
19 units
Distribution
Florida
Code information
UDI/DI number 00887761984622
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 887761984622