Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1650-2026
- Category
- Medical devices
- Recall date
- 2026-02-10
- Enforcement report date
- 2026-04-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-04-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1650-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"
- Hazard / reason
- Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
- Quantity
- 106,621 US; 2,350 OUS
- Distribution
- Domestic distribution nationwide US. International distribution to Mexico and Panama.
- Code information
- MSC8522 - Puracol UDI-DI 10080196285860 (each) 40080196285861 (case) Lot P2090 MSC8622EP - Puracol Plus UDI-DI 10080196296804 (each) 40080196296805 (case) Lot PL2107
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10080196285860
- 10080196296804