BD Kiestra" ReadA; Catalog No.: 446948.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1658-2026
- Category
- Medical devices
- Recall date
- 2026-02-06
- Enforcement report date
- 2026-04-08
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- BD KIESTRA LAB AUTOMATION
From the FDA Enforcement Report dated 2026-04-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1658-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- BD Kiestra" ReadA; Catalog No.: 446948.
- Hazard / reason
- In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.
- Quantity
- 82 units (7 US, 65 OUS)
- Distribution
- Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;
- Code information
- Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 382904469481