ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1690-2026
- Category
- Medical devices
- Recall date
- 2026-02-13
- Enforcement report date
- 2026-04-08
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-04-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1690-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
- Hazard / reason
- Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
- Quantity
- 651,789
- Distribution
- Worldwide - US Nationwide and the countries of Canada.
- Code information
- UDI-DI 10888277314719 (ea) 40888277314710(case) All Lots
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10888277314719