Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1706-2026
- Category
- Medical devices
- Recall date
- 2026-02-16
- Enforcement report date
- 2026-04-08
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-04-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1706-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
- Hazard / reason
- Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- Quantity
- 14,379 kits total
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
- Code information
- 1. UDI-DI 10198459629334 (EA) 40198459629335 (CS) LOTS 26AMG895 2. UDI-DI 10198459545283 (EA) 40198459545284 (CS) LOTS 25LBJ366 26ABR365 3. UDI-DI 10195327471859 (EA) 40195327471850 (CS) LOTS 24EBA317 24EBV431 24IBD694 24JBQ783
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10198459629334
- 10198459545283
- 10195327471859