NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1713-2026
- Category
- Medical devices
- Recall date
- 2026-02-27
- Enforcement report date
- 2026-04-15
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-04-15. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1713-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.
- Hazard / reason
- Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
- Quantity
- 192690 units
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
- Code information
- 1. Medline SKU 70083007, UDI/DI each 10193489065855, UDI/DI case 30193489065859, Lot Numbers: 0000077029, 0000081537, 0000084335, 0000087347, 0000089563, 0000113792, 0000115753, 0000122956, 0000127785, 0000131986, 0000158748, 0000175535, 0000187926; 2. Medline SKU 70084007, UDI/DI each 10193489065893, UDI/DI case 30193489065897, Lot numbers: 0000078186, 0000081539, 0000084274, 0000086756, 0000089554, 0000094842, 0000096918, 0000105613, 0000109479, 0000111244, 0000114562, 0000115421, 0000117317, 0000118251, 0000120265, 0000120769, 0000121714, 0000123466, 0000124589, 0000131984, 0000131989, 0000138071, 0000140322, 0000142850, 0000147002, 0000152069, 0000162694, 0000173507, 0000174662, 0000174773, 0000176555, 0000178806, 0000189895; 3. Medline SKU 70084017, UDI/DI each 10193489065916, UDI/DI case 20193489065913, Lot Numbers: 0000084509, 0000086823, 0000095687, 0000165551, 0000165868, 0000166300; 4. Medline SKU 80083007, UDI/DI each 10193489066753, UDI/DI case 30193489066757, Lot Numbers: 0000097388, 0000127072, 0000135012, 0000137004; 5. Medline SKU 80084007, UDI/DI each 10193489066791, UDI/DI case 30193489066795, Lot Numbers: 0000080553, 0000080919, 0000082969, 0000085041, 0000088044, 0000094514, 0000099870, 0000101882, 0000102372, 0000105155, 0000107328, 0000109442, 0000112810, 0000112966, 0000114317, 0000115744, 0000117572, 0000118722, 0000118962, 0000119950, 0000120931, 0000121305, 0000121393, 0000122583, 0000122606, 0000123102, 0000126198, 0000126566, 0000127132,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I