MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)

Agency
U.S. Food and Drug Administration
Recall number
Z-1740-2026
Category
Medical devices
Recall date
2026-02-13
Enforcement report date
2026-04-08
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic MiniMed, Inc.

From the FDA Enforcement Report dated 2026-04-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Quantity
60,880
Distribution
Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
Code information
Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 630G Insulin Pump: 763000253066, 643169752726, 763000101527, 763000190446, 763000317591, 763000157043, 643169752726, 763000101527, 763000190446, 763000317591, 643169873841, 643169878648, 763000179632, 76300031759101, 76300031759103, 76300031762103, 763000367091, 763000317621, 763000157043, 76300031762101: MiniMed 630G System(M020092C004 1); MINIMED 630G and MINIMED 670G INSULIN PUMP SYSTEM(M042888C002 1); MiniMed 630G Insulin pump MODEL MMT-1714(M993428A262 1); MiniMed 630G System User Guide(M020092C001 3, M020092C003 1, M020092C005 1); MINIMED 630G SYSTEM USER GUIDE mg/dL(M979512A011 1); MINIMED 630G WITH SMARTGUARD TECHNOLOGY SYSTEM USER GUIDE(MP6026120-262 1); MINIMED 630G WITH SMARTGUARD TECHNOLOGY SYSTEM USER GUIDE(M030278C001 3, M030278C002 1); MINIMED 630G WITH SMARTGUARD TECHNOLOGY SYSTEM USER GUIDE mmol/L (MP6026120-024 1); MiniMed Veo with Enlite Threshold Suspend Performance(M10577668A262 1).
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Model numbers
  • MMT-1714

Brands