Burlington Medical, Kilt Blockers.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1765-2026
- Category
- Medical devices
- Recall date
- 2026-02-11
- Enforcement report date
- 2026-04-08
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Burlington Medical, LLC
From the FDA Enforcement Report dated 2026-04-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1765-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Potential for attenuation degradation over time, decreasing the lifespan.
- Quantity
- 94 units
- Distribution
- Worldwide distribution. Nationwide distribution in the US and international distribution to Canada, Australia, Bangladesh, Chile, Hong Kong, India, Mexico, Malaysia, Panama, Singapore, Thailand, Taiwan, Uruguay, and Vietnam.
- Code information
- All serial numbers are affected that were manufactured with Xenolite 800 attenuation material from 01/16/2025 to 01/21/2026. 1. Model Number: R8KILTBLOCKER10X12; UDI-DI (Product Code): 00840331272727 (KILTBLOCKER8-R8KILTBLOCKER10X12). 2. Model Number: R8KILTBLOCKER12X12; UDI-DI (Product Code): 00197513011293 (KILTBLOCKER8-R8KILTBLOCKER12X12). 3. Model Number: R8KILTBLOCKER18X20; UDI-DI (Product Code): 00840331272741 (KILTBLOCKER8-R8KILTBLOCKER18X20). 4. Model Number: R8KILTBLOCKER20X24; UDI-DI (Product Code): 00840331272758 (KILTBLOCKER8-R8KILTBLOCKER20X24). 5. Model Number: R8KILTBLOCKER24X24; UDI-DI (Product Code): 00840331272765 (KILTBLOCKER8-R8KILTBLOCKER24X24). 6. Model Number: S8KILTBLOCKER10X10; UDI-DI (Product Code): 00840331272772 (KILTBLOCKER8-S8KILTBLOCKER10X10). 7. Model Number: S8KILTBLOCKER10X12; UDI-DI (Product Code): 00840331272789 (KILTBLOCKER8-S8KILTBLOCKER10X12). 8. Model Number: S8KILTBLOCKER12X12; UDI-DI (Product Code): 00197513011309 (KILTBLOCKER8-S8KILTBLOCKER12X12). 9. Model Number: S8KILTBLOCKER18X20; UDI-DI (Product Code): 00840331272802 (KILTBLOCKER8-S8KILTBLOCKER18X20). 10. Model Number: S8KILTBLOCKER20X24; UDI-DI (Product Code): 00840331272819 (KILTBLOCKER8-S8KILTBLOCKER20X24). 11. Model Number: S8KILTBLOCKER24X24; UDI-DI (Product Code): 00840331272826 (KILTBLOCKER8-S8KILTBLOCKER24X24). 12. Model Number: U8KILTBLOCKER12X12; UDI-DI (Product Code): 00197513011316 (KILTBLOCKER8-U8KILTBLOCKER12X12). 13. Model Number:…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Model numbers
- R8KILTBLOCKER10X12
- R8KILTBLOCKER12X12
- R8KILTBLOCKER18X20
- R8KILTBLOCKER20X24
- R8KILTBLOCKER24X24
- S8KILTBLOCKER10X10
- S8KILTBLOCKER10X12
- S8KILTBLOCKER12X12
- S8KILTBLOCKER18X20
- S8KILTBLOCKER20X24
- S8KILTBLOCKER24X24
- U8KILTBLOCKER12X12
- U8KILTBLOCKER18X20
- U8KILTBLOCKER20X24
- U8KILTBLOCKER24X24