B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1795-2026
- Category
- Medical devices
- Recall date
- 2026-03-12
- Enforcement report date
- 2026-04-15
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Brahms GmbH
From the FDA Enforcement Report dated 2026-04-15. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1795-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor
- Hazard / reason
- Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
- Quantity
- 5,524
- Distribution
- Worldwide distribution. US states: NC, TX, MN, UT, NY, FL, OH, IL, CT, CA, GA. Other countries: DE, AT, AL, AU, BG, BR, CA, CH, CZ, DK, EC, ET, ES, FR, GB, GR, HK, HU, ID, IL, IT, KW, KZ, LT, MK, ML, MX, NL, NO, PA, PL, PT, RO, RS, SE, SI, TH, TU, TW, UA, ZA
- Code information
- REF: 89986, UDI-DI: 04260157638976, Lots: 40102+R78, 40102+R79, 40102+R80. REF: 859075N, UDI-DI: 04260157632196, All Lots.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 4260157638976
- 4260157632196