The product is a handheld ultraviolet-C germicidal wand and contains two low-pressure mercury bulbs with a peak…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1807-2026
- Category
- Medical devices
- Recall date
- 2026-03-06
- Enforcement report date
- 2026-04-29
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Uvlizer c/o RAIS INTERNATIONAL LLC
From the FDA Enforcement Report dated 2026-04-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1807-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- The product is a handheld ultraviolet-C germicidal wand and contains two low-pressure mercury bulbs with a peak emission wavelength at 254 nm.
- Hazard / reason
- In some foreseeable use conditions, the Products can expose nearby persons to UVC radiation at levels significantly above limits recommended by international safety guidelines for skin and eye exposure.
- Quantity
- 480,000
- Distribution
- U.S.
- Code information
- None provided.
- Recall type
- FDA Mandated
- Product type
- Devices
- Classification (FDA wording)
- Class II