ARTIS Pheno VE30A and VE40A, Model 10849000
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1814-2026
- Category
- Medical devices
- Recall date
- 2026-03-12
- Enforcement report date
- 2026-04-22
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Siemens Medical Solutions USA, Inc
From the FDA Enforcement Report dated 2026-04-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1814-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
- Quantity
- 73
- Distribution
- U.S. and O.U.S.
- Code information
- UDI 04056869046877
- Recall type
- FDA Mandated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 4056869046877
- Possible model numbers (as listed; not verified as model numbers)
- 10849000