Brand Name: Elegance¿ Anterior Cervical Plate Product Name: ELEGANCE 1 LEVEL CERVICAL PLATE, ELEGANCE 2 LEVEL CERVICAL…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1821-2026
- Category
- Medical devices
- Recall date
- 2026-02-20
- Enforcement report date
- 2026-04-22
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Clariance-SAS
From the FDA Enforcement Report dated 2026-04-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1821-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Elegance¿ Anterior Cervical Plate Product Name: ELEGANCE 1 LEVEL CERVICAL PLATE, ELEGANCE 2 LEVEL CERVICAL PLATE, ELEGANCE 3 LEVEL CERVICAL PLATE, ELEGANCE 4 LEVEL CERVICAL PLATE Model/Catalog Number: See Excel sheet attached named "FDA 5072 Product Information" Software Version: Not applicable Product Description: The Elegance¿Anterior Cervical plate is designed for use as a cervical plate system. It has a shape which restores the intervertebral height and lordosis. The Elegance¿ Anterior Cervical plate consists in a variety of shapes and sizes of cervical plates with rounded corners, featuring bone screw s holes, locking systems and bone screws. The plates go from one (1) to five (5) levels, and the screws are selfdrilling or self- tapping, and fixed or variable. The fixation is provided by the bone screws inserted into the vertebral body by means of an anterior approach. The Elegance¿plates, as well as the bone screws, are made of compliant ASTM F136 Titanium alloy. It is essential to insert the implants with the instrumentation specifically designed for this purpose. For more description of the instrumentation, read the technical documentation associated to the Elegance¿ Anterior Cervical plate. Detailed information concerning the surgical technique of the Elegance¿Anterior Cervical plate is available upon request, please contact CLARIANCE or its local representative.
- Hazard / reason
- Due to a quality issue regarding the presence of burr on the finished cervical plate product.
- Quantity
- 511
- Distribution
- U.S. Nationwide distribution in the states of CA, IL, NY and TX.
- Code information
- Lot Code: See Excel sheet attached named "FDA 5072 Product Information" See Spreadsheet attached named "FDA 5072 Product Quantity Distributed" Catalog Number: 53430108 (ELEGANCE 1 LEVEL CERVICAL PLATE 8MM) UDI-DI code: 03700780644353 Lot Numbers: PC03X R879X Catalog Number: 53430110 (ELEGANCE 1 LEVEL CERVICAL PLATE 10MM) UDI-DI code: 03700780644353 Lot Numbers: P758X RAA2X R825X R902X Catalog Number: 53430111 (ELEGANCE 1 LEVEL CERVICAL PLATE 11MM) UDI-DI code: 03700780644384 Lot Numbers: R194X RAA1X Catalog Number: 53430112 (ELEGANCE 1 LEVEL CERVICAL PLATE 12MM) UDI-DI code: 03700780644391 Lot Numbers: R195X RAB2X Catalog Number: 53430113 (ELEGANCE 1 LEVEL CERVICAL PLATE 13MM) UDI-DI code: 03700780644407 Lot Numbers: R196X R197X R584X RA81X Catalog Number: 53430114 (ELEGANCE 1 LEVEL CERVICAL PLATE 14MM) UDI-DI code: 03700780644414 Lot Number: R585X Catalog Number: 53430116 (ELEGANCE 1 LEVEL CERVICAL PLATE 16MM) UDI-DI code: 03700780644438 Lot Number: R586X Catalog Number: 53430120 (ELEGANCE 1 LEVEL CERVICAL PLATE 20MM) UDI-DI code: 03700780644452 Lot Number: R182X Catalog Number: 53430220 (ELEGANCE 2 LEVEL CERVICAL PLATE 20MM) UDI-DI code: 03700780644452 Lot Numbers: R587X RAA0X Catalog Number: 53430224 (ELEGANCE 2 LEVEL CERVICAL PLATE 24MM) UDI-DI code: 03700780644490 Lot Number: R588X Catalog Number: 53430228 (ELEGANCE 2 LEVEL CERVICAL PLATE 28MM) UDI-DI code: 03700780644513 Lot Numbers: R589X RA80X Catalog Number: 53430230 (ELEGANCE 2 LEVEL CERVICAL PLATE…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 3700780644353
- 3700780644384
- 3700780644391
- 3700780644407
- 3700780644414
- 3700780644438
- 3700780644452
- 3700780644490
- 3700780644513
- 3700780644520
- 3700780644537
- 3700780644544
- 3700780644551
- 3700780644575
- 3700780644582
- 3700780644605
- 3700780644612
- 3700780644629
- 3700780644636
- 3700780644643
- 3700780644667