Philips Spectral CT on Rails. Model Number: 728334.

Agency
U.S. Food and Drug Administration
Recall number
Z-1825-2026
Category
Medical devices
Recall date
2026-03-07
Enforcement report date
2026-04-22
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
PHILIPS MEDICAL SYSTEMS

From the FDA Enforcement Report dated 2026-04-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Quantity
3 systems
Distribution
Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
Code information
Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Possible model numbers (as listed; not verified as model numbers)
  • 728334