Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EXTREMITY PACK DYNJ34846C HAND…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1887-2026
Category
Medical devices
Recall date
2026-02-27
Enforcement report date
2026-04-29
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-04-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EXTREMITY PACK DYNJ34846C HAND PACK DYNJ83218A, DYNJ85396 LAMINECTOMY DYNJ905156R MAJOR EXTREMITY DYNJ905159L, DYNJ905159M MAJOR EXTREMITY PACK DYNJ69926A PODIATRY PACK DYNJ41250B
Hazard / reason
Unapproved design changes to the products outside of the 510(k) clearance.
Quantity
270,311 total
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Code information
DYNJ34846C UDI-DI 10193489781441 (EA) 40193489781442 (CS) LOT 24BBA459 DYNJ83218A UDI-DI 10195327382308 (EA) 40195327382309 (CS) LOT 24BBD199 DYNJ85396 UDI-DI 10195327497613 (EA) 40195327497614 (CS) LOT 24BBA503 DYNJ905156R UDI-DI 10198459017704 (EA) 40198459017705 (CS) LOTS 24HBG726 24JBJ484 25ABS758 25BBR196 25CBN174 25EBE136 DYNJ905159L UDI-DI 10195327664251 (EA) 40195327664252 (CS) LOT 24EMB726 DYNJ905159M UDI-DI 10198459017735 (EA) 40198459017736 (CS) LOT 24GMA714 24HMD401 24HMJ945 24JMD201 24KMG592 25AMC024 25BMD945 25CMG373 DYNJ69926A UDI-DI 10195327153601 (EA) 40195327153602 (CS) LOT 24ABS445 DYNJ41250B UDI-DI 10888277801257 (EA) 40888277801258 (CS) LOT 24ABO122 ***Updated 5/20/26 - The following lots were included in the customer letter but not the FDA submission.*** DYNJ905156R lots 24GBD135 24IBK563 24IBS862 25BBH767
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10193489781441
  • 10195327382308
  • 10195327497613
  • 10198459017704
  • 10195327664251
  • 10198459017735
  • 10195327153601
  • 10888277801257

Brands