Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

Agency
U.S. Food and Drug Administration
Recall number
Z-1923-2026
Category
Medical devices
Recall date
2026-03-25
Enforcement report date
2026-04-29
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Remel Europe Ltd.

From the FDA Enforcement Report dated 2026-04-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Hazard / reason
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
Quantity
117 US; 950 OUS
Distribution
Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
Code information
UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 5056080500270

Brands