Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1923-2026
- Category
- Medical devices
- Recall date
- 2026-03-25
- Enforcement report date
- 2026-04-29
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Remel Europe Ltd.
From the FDA Enforcement Report dated 2026-04-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1923-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
- Hazard / reason
- Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
- Quantity
- 117 US; 950 OUS
- Distribution
- Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.
- Code information
- UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 5056080500270