Sequencing Agent SEQ0067, component of FoundationOne CDx (F1CDx)

Agency
U.S. Food and Drug Administration
Recall number
Z-1927-2026
Category
Medical devices
Recall date
2026-03-20
Enforcement report date
2026-04-29
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Foundation Medicine, Inc.

From the FDA Enforcement Report dated 2026-04-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants
Quantity
163 units
Distribution
US Nationwide distribution in the states of Massachusetts and North Carolina.
Code information
Model Number: SEQ0067. Lot Numbers: 24346, 24386, 24488.