iLet Bionic Pancreas, REF: BB1001

Agency
U.S. Food and Drug Administration
Recall number
Z-1996-2026
Category
Medical devices
Recall date
2026-03-12
Enforcement report date
2026-05-06
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Beta Bionics, Inc.

From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
Quantity
1,080
Distribution
US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Code information
Software versions 1.4.3, and 1.4.2
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II