Heraeus, PALAMIX duo. Material Number: 66057897.

Agency
U.S. Food and Drug Administration
Recall number
Z-1998-2026
Category
Medical devices
Recall date
2026-03-25
Enforcement report date
2026-05-06
Classification
Class III
Status as reported by FDA
Ongoing
Manufacturer / firm
Heraeus Medical GmbH (Dental Division)

From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Quantity
1,125 units
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Code information
Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)