Brand Name: Leksell Vantage Arc System REF: 1053958
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2006-2026
- Category
- Medical devices
- Recall date
- 2026-03-19
- Enforcement report date
- 2026-05-06
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Elekta, Inc.
From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2006-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Leksell Vantage Arc System REF: 1053958
- Hazard / reason
- Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
- Quantity
- 768 (364Left/364Right)
- Distribution
- US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea, Republic of Kuwait Lithuania Mexico Morocco Netherlands New Zealand Norway Pakistan Panama Poland Portugal Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Kingdom
- Code information
- Lot Code: Arc System: UDI:07340048304856 Parts of Arc System: Left Z-Slide: Serial Numbers :UDI: 07340048304818 00001 00625 Right Z-Slide: UDI: 07340048304801/ Serial Numbers: 00001 00675
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 7340048304856
- 7340048304818
- 7340048304801