Brand Name: Leksell Vantage Arc System REF: 1053958

Agency
U.S. Food and Drug Administration
Recall number
Z-2006-2026
Category
Medical devices
Recall date
2026-03-19
Enforcement report date
2026-05-06
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Elekta, Inc.

From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Leksell Vantage Arc System REF: 1053958
Hazard / reason
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
Quantity
768 (364Left/364Right)
Distribution
US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea, Republic of Kuwait Lithuania Mexico Morocco Netherlands New Zealand Norway Pakistan Panama Poland Portugal Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Kingdom
Code information
Lot Code: Arc System: UDI:07340048304856 Parts of Arc System: Left Z-Slide: Serial Numbers :UDI: 07340048304818 00001 00625 Right Z-Slide: UDI: 07340048304801/ Serial Numbers: 00001 00675
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 7340048304856
  • 7340048304818
  • 7340048304801

Brands