Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF, Medline Kit…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2019-2026
Category
Medical devices
Recall date
2026-03-19
Enforcement report date
2026-05-06
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920D; 2. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920F; 3. MEM MAXILLOFACIAL PACK-LF, Medline Kit Number/SKU DYNJ0660040S; 4. ORAL SURGERY PACK, Medline Kit Number/SKU DYNJ66450B; 5. ORAL SURGERY PACK RFID, Medline Kit Number/SKU DYNJ66450C; 6. ORAL SURGERY PACK, Medline Kit Number/SKU DYNJ66450D; 7. ORAL SURGERY PACK RFID-V2, Medline Kit Number/SKU DYNJ66450F; 8. ORAL SURGERY PACK RFID-V2, Medline Kit Number/SKU DYNJ66450G.
Hazard / reason
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Quantity
2228 kits
Distribution
Worldwide distribution.
Code information
Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23JMB165; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23CMH103; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23BMH200; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23BMB495; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 22KMH108; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 22KME285; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23KMJ648; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 26BME013; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25JMC725; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25IMA571; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25HMC497; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25FMG199; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714,…