Medline medical procedure kits, containing Medline Neuro Sponges (102 in total)
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2025-2026
- Category
- Medical devices
- Recall date
- 2026-03-19
- Enforcement report date
- 2026-05-06
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2025-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
- Quantity
- 22116 kits
- Distribution
- Worldwide distribution.
- Code information
- Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23KBN711; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23GBD593; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23GBD781; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23DBF906; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22LBI985; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22KBC504; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22IBF741; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22FBX069; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22DBS948; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22CBX667; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 21IBT137; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 21HBG608; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 25GBV215;…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II