Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2026-2026
Category
Medical devices
Recall date
2026-03-19
Enforcement report date
2026-05-06
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-05-06. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060J; 5. NEURO BASIN KIT, Medline Kit Number/SKU DYNJ80200B; 6. NEURO-SPINE BASIN PACK, Medline Kit Number/SKU DYNJ83019; 7. ENT PACK, Medline Kit Number/SKU DYNJ86328; 8. EDMOND SPINE PACK, Medline Kit Number/SKU DYNJ89263; 9. ENT PACK, Medline Kit Number/SKU DYNJ89675; 10. ENT PACK, Medline Kit Number/SKU DYNJ89675A; 11. ENT/PAIN PACK, Medline Kit Number/SKU DYNJT7493.
Hazard / reason
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Quantity
9805 kits
Distribution
Worldwide distribution.
Code information
Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25IMD003; Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25HMH643; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22GBS127; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22FBB472; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25FBE752; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25EBA260; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25DBG176; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25CBD819; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25BBA593; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 24KBJ163; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23KMI415; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMK023; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMB211;…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands