LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2048-2026
- Category
- Medical devices
- Recall date
- 2026-03-30
- Enforcement report date
- 2026-05-13
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Boston Scientific Corporation
From the FDA Enforcement Report dated 2026-05-13. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2048-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
- Quantity
- 2 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Belgium.
- Code information
- LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 802526620713
- Model numbers