See RES for complete list. Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2098-2026
- Category
- Medical devices
- Recall date
- 2026-01-07
- Enforcement report date
- 2026-05-13
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-05-13. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2098-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- See RES for complete list. Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W; 4) MAJOR LAPAROTOMY CDS, Model Number: CDS860015X; 5) MINOR LAPAROTOMY CDS, Model Number: CDS860016P; 6) MINOR LAPAROTOMY CDS, Model Number: CDS860016Q; 7) MINOR LAPAROTOMY CDS, Model Number: CDS860016R; 8) MINOR LAPAROTOMY CDS, Model Number: CDS860016S; 9) LAP GASTRIC BYPASS CDS, Model Number: CDS860146M;
- Hazard / reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Quantity
- 114778 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- 1) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21FBR386; 2) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21GBS688; 3) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21IBL031; 4) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21KBB595; 5) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21LBD726; 6) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21LBT172; 7) CDS860015V, UDI-DI: 10195327209582(each), 40195327209583(case), Lot Number: 22HBF170; 8) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23FBP597; 9) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23GBM065; 10) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23HBT489; 11) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23JBR469; 12) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23KBX194; 13) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23LBQ388; 14) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), Lot Number: 24DBC504; 15) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), Lot Number: 24DBP400; 16) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), Lot Number: 24EBA340; 17) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case),…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10193489847994
- 10195327209582
- 10195327349509
- 10195327647513
- 10193489848007
- 10195327209599
- 10195327349479
- 10195327647520
- 10195327224547
- Model numbers
- CDS860015U
- CDS860015V
- CDS860015W
- CDS860015X
- CDS860016P
- CDS860016Q
- CDS860016R
- CDS860016S
- CDS860146M