See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2114-2026
- Category
- Medical devices
- Recall date
- 2026-01-07
- Enforcement report date
- 2026-05-13
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-05-13. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2114-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W;
- Hazard / reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Quantity
- 160980 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- 1) CDS940087AG, UDI-DI: 10193489847260(each), 40193489847261(case), Lot Number: 21EBC900; 2) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21GBI183; 3) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21IBJ260; 4) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21JBF936; 5) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21KBH163; 6) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22BBY060; 7) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBA274; 8) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBF695; 9) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBG548; 10) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22GBE442; 11) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21BBL223; 12) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21FBS254; 13) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21GBR895; 14) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBU197; 15) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBV695; 16) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBB625; 17) CDS980832R, UDI-DI: 10193489900972(each),…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10193489847260
- 10193489933758
- 10193489788655
- 10193489900972
- 10195327064495
- 10193489900989
- 10195327064501
- 10193489304206
- 10193489929157
- Model numbers
- CDS940087AG
- CDS940087AI
- CDS980832Q
- CDS980832R
- CDS980832S
- CDS980833Q
- CDS980833R
- CDS980865V
- CDS980865W