Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2126-2026
- Category
- Medical devices
- Recall date
- 2026-01-07
- Enforcement report date
- 2026-05-13
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-05-13. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2126-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189A
- Hazard / reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Quantity
- 59 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- 1) DYKMBNDL106, UDI-DI: 10193489479294(each), 40193489479295(case), Lot Number: 21ABJ542; 2) DYKMBNDL106, UDI-DI: 10193489479294(each), 40193489479295(case), Lot Number: 21BBG085; 3) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 21KLA631; 4) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22ALA713; 5) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22BLB059; 6) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22CLA049; 7) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21ABS539; 8) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21ABS540; 9) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21CBS043; 10) DYKMBNDL189, UDI-DI: 10193489478457(each), 40193489478458(case), Lot Number: 21ABI996; 11) DYKMBNDL189, UDI-DI: 10193489478457(each), 40193489478458(case), Lot Number: 21BBG092; 12) DYKMBNDL189A, UDI-DI: 10193489880823(each), 40193489880824(case), Lot Number: 21JLA521
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10193489479294
- 10193489880816
- 10193489478549
- 10193489478457
- 10193489880823
- Model numbers
- DYKMBNDL106
- DYKMBNDL106A
- DYKMBNDL188
- DYKMBNDL189
- DYKMBNDL189A