Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2156-2026
- Category
- Medical devices
- Recall date
- 2026-04-10
- Enforcement report date
- 2026-05-20
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ARROW INTERNATIONAL, LLC
From the FDA Enforcement Report dated 2026-05-20. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2156-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
- Quantity
- 285
- Distribution
- Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
- Code information
- REF/UDI-DI/Lot: CS-15242-I/ 30801902195280/33F23G0536; CS-15282-I/30801902195532/33F23C0877, 33F24A0281, 33F24H0041; CS-15322-I/30801902197451/33F23G0545
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I