Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2157-2026
Category
Medical devices
Recall date
2026-04-10
Enforcement report date
2026-05-20
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
ARROW INTERNATIONAL, LLC

From the FDA Enforcement Report dated 2026-05-20. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15312-VFIE, CS-15422-VFI, CS-15422-VFIE, CS-15552-VFI, CS-15552-VFIE; Arrow-Clark VectorFlow Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: REF: CS-15192-VFIM, CS-15232-VFIM, CS-15272-VFIM; Arrow-Clark VectorFlow Retrograde Chronic Hemodialysis Catheter, REF: ACS-15192-VF, ACS-15232-VF, ACS-15272-VF, ACS-15552-VF, CS-15192-VF, CS-15192-VFE, CS-15232-VF, CS-15232-VFE, CS-15272-VF, CS-15272-VFE, CS-15312-VF, CS-15312-VFE, CS-15422-VF, CS-15422-VFE, CS-15552-VF, CS-15552-VFE; Arrow-Clark VectorFlow Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: CS-15192-VFM, CS-15232-VFM, CS-15272-VFM
Hazard / reason
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
Quantity
92,827
Distribution
Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
Code information
REF/UDI-DI/Lots: ACS-15192-VFI/ 10801902126709/33F23G0040, 33F25C0689; ACS-15232-VFI/10801902126723/33F23G0045, 33F25C0678; ACS-15272-VFI/ 10801902126747/33F24G0638; CS-15192-VFI/10801902127188, 10801902193572/33F23E0579, 33F23G0042, 33F23K0170, 33F23L0933, 33F24E0271, 33F24G0166, 33F24J0086, 33F24L0064, 33F24M0224, 33F25C0652, 33F25D0052, 33F25D0700, 33F25E1081, 33F25F0015, 33F25F0388, 33F25G0702, 33F25H0785; CS-15192-VFIE/10801902199659/33F23D0853, 33F23E0773, 33F23G0044, 33F23G0532, 33F23H0611, 33F23K0169,33F23K1105,33F23L0934, 33F24A0912, 33F24A0913, 33F24C0058, 33F24D1001, 33F24G0167,33F24G0187, 33F24J0030, 33F24J0032, 33F24J0339, 33F24J0381, 33F24K0659, 33F24L0058, 33F24L0511, 33F24M0155, 33F24M0215, 33F24M0265, 33F25A0143, 33F25A0407, 33F25B0259, 33F25C0037, 33F25D0285, 33F25D0738, 33F25F0060, 33F25G0635; CS-15232-VFI/10801902194463/33F23C0864, 33F23E0578, 33F23E0832, 33F23G0535, 33F23G0840, 33F23H0612, 33F23K0144, 33F24A0914, 33F24C0079, 33F24E0269, 33F24J0042, 33F25B0256, 33F25C0014, 33F25C0709, 33F25C0650, 33F25E1085, 33F25E1130, 33F25F0009, 33F25G0023, 33F25G0450, 33F25H0341, 33F25J1014; CS-15232-VFIE/10801902195149/33F23D0855, 33F23E0774, 33F23G0047, 33F23G0048, 33F23G0049, 33F23K0145, 33F23K0175, 33F23L0938, 33F24A0915, 33F24C0069, 33F24G0169, 33F24J0026, 33F24J0027, 33F24J0028, 33F24J0084, 33F24K0700, 33F24L0178, 33F24M0005, 33F24M0214, 33F24M0219, 33F25C0646, 33F25C0665, 33F25D0097, 33F25E0358, 33F25E0434, 33F25F0012, 33F25F0275, 33F25F0328, 33F25G0069,…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Brands